• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL 4.75MM HEALIX ADVANCE KNOTLESS BR ANCHOR; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL 4.75MM HEALIX ADVANCE KNOTLESS BR ANCHOR; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 222885
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/10/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi:(b)(4).E3: reporter is a j&j employee.H4: the device manufacture date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported by the affiliate in japan that during an arthroscopic rotator cuff repair procedure on (b)(6) 2023, it was observed that the 4.75mm healix advance knotless br anchor device was damaged in spiral shape; and therefore, could not be used.According to the report, the event was noticed when the anchor was inserted into the greater tubercle for lateral bridging.Another like device was used to complete the procedure successfully within 30 minutes delay.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: h4: the device manufacture date was reported as unknown on the initial report; and has been updated accordingly.Investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information or the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number 116l891, and no non-conformance was identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.75MM HEALIX ADVANCE KNOTLESS BR ANCHOR
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key17406554
MDR Text Key320663343
Report Number1221934-2023-02823
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number222885
Device Lot Number116L891
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-