It was reported that the stent was difficult to advance, difficult to remove, and partially deployed.This 6 x 40 x 130 innova vascular stent was selected for use in an arterial stent procedure for superficial femoral artery (sfa) stenosis.During the procedure, a non-boston scientific guidewire was used as a buddy wire in the sheath with the stent system and non-boston scientific guidewire.It was noted there was not enough room for both devices in the sheath and difficulties advancing occurred.Additionally, difficulties were encountered when removing the stent.While removing, the stent partially deployed at the sheath valve.The stent was fully removed manually, and the procedure was completed using the standard technique without the use of a buddy wire and another of the same device.There were no patient complications.
|