• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number 26924
Device Problems Premature Activation (1484); Difficult to Remove (1528); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2023
Event Type  malfunction  
Event Description
It was reported that the stent was difficult to advance, difficult to remove, and partially deployed.This 6 x 40 x 130 innova vascular stent was selected for use in an arterial stent procedure for superficial femoral artery (sfa) stenosis.During the procedure, a non-boston scientific guidewire was used as a buddy wire in the sheath with the stent system and non-boston scientific guidewire.It was noted there was not enough room for both devices in the sheath and difficulties advancing occurred.Additionally, difficulties were encountered when removing the stent.While removing, the stent partially deployed at the sheath valve.The stent was fully removed manually, and the procedure was completed using the standard technique without the use of a buddy wire and another of the same device.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17408299
MDR Text Key320069685
Report Number2124215-2023-39922
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26924
Device Catalogue Number26924
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/25/2023
Initial Date FDA Received07/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-