• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR Back to Search Results
Model Number 71005
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Inadequate Pain Relief (2388)
Event Date 06/29/2023
Event Type  Injury  
Event Description
Patient was implanted with the nalu spinal cord stimulator system on (b)(6) 2023 and subsequently reported difficulty with communication between the implantable pulse generator (ipg) and the external therapy discs.The ipg depth was measured and found to be deeper than the optimal recommendation for communication.Surgical revision was performed on (b)(6) 2023 to implant a replacement ipg in a more superficial location.
 
Manufacturer Narrative
There are no allegations of failure or malfunction of the nalu system or its components.Communication between the ipg and therapy discs is directly related to the depth in which the ipg was implanted.Patient reports issues are resolved after placing the ipg at an appropriate depth.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008 2377
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad CA 92008 7241
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key17413002
MDR Text Key319933690
Report Number3015425075-2023-00167
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00812537036137
UDI-Public01008125370361371122102117251021
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71005
Device Catalogue Number71005
Device Lot NumberA05190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-