MAUDE Adverse Event Report: SYNTHES GMBH BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.715
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the battery reciprocator device was leaking fluid.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi (b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the battery reciprocator device had fluid ingress and would not run.It was further determined that the device failed pretest for check the function of the device and check the oscillation frequency with frequency meter.Therefore, the reported condition that the device was leaking fluid was confirmed.The assignable root cause of this condition was determined to be traced to maintenance, which is improper maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY RECIPROCATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key17413784
MDR Text Key319942754
Report Number8030965-2023-09476
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819491854
UDI-Public07611819491854
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/17/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown