Model Number 404683
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/04/2023
Event Type  malfunction  
Event Description
It was reported that the device was explanted due to eri status approx.66 months after the implantation.No adverse patient events were reported.Should additional information be received, this file will be update.
 
Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.There was no indication of a material or manufacturing problem.Subsequently the device was interrogated.The interrogation could be properly performed and revealed the mos2 battery status.All electrical parameters were verified and showed a normal device behavior.Especially, the current consumption and the battery state of discharge were found normal and as expected.There was no indication of a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTICA 5 HF-T DF-1 IS-1 PROMRI
Type of Device
NO MATCH
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key17415349
MDR Text Key320119744
Report Number1028232-2023-03793
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
PMA/PMN Number
P050023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model Number404683
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-