Catalog Number D134721IL
Device Problems Erratic or Intermittent Display (1182); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2023
Event Type  malfunction  
Manufacturer Narrative
The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.A visual inspection and magnetic sensor functionality test of the returned device were performed following bwi procedures.Visual analysis revealed foreign material and a hole in the surface of the pebax.The device was connected to the carto 3 system and it was recognized and visualized correctly.No icon issues were observed.The failure observed in the pebax could be related to the wrong manipulation of the device during the procedure and to the issue reported by the customer; however, this can not be conclusively determined.A manufacturing record evaluation was performed for the finished device 30959009l number, and no internal actions related to the reported complaint condition were identified.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use contain the following recommendations: in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and the bwi product analysis lab identified a hole on the surface of the pebax.It was reported that during the operation, the icon was waggling on the carto system screen.A second device was used to complete the operation.There was no adverse event reported on patient.Icon jumping is not mdr-reportable.Hole in the pebax is mdr-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17422400
MDR Text Key320021141
Report Number2029046-2023-01637
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2023
Device Catalogue NumberD134721IL
Device Lot Number30959009L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2023
Date Manufacturer Received07/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-