It was reported that during preparation, the hemo valve rotating was connected to the catheter; however, after applying negative pressure it was noted air and the o-ring not sealing.There was no reported device use or patient involvement, and there was no clinically significant delay in the procedure.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It is possible that the device was not completely tightened/connected resulting in the reported leak; however as the device was not returned for analysis this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|