• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C37101329-NLJ
Device Problem Gas/Air Leak (2946)
Patient Problem Insufficient Information (4580)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
It was reported that a leak alarm occurred and use was discontinued.Patient involvement unknown.
 
Manufacturer Narrative
Other, month and year of event have been provided, day is unknown.Udi number is unknown, no information has been provided to date.510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other text: h3, h6: event methods, evaluation, and conclusion codes: updated.Four photos of the device and one sample were received for investigation.The reported issue was confirmed during visual inspection, which confirmed there was a hole in the device breathing bag.A review of the manufacturing device history records (dhr) shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with the reported lot of products.The investigation attributed the root cause of this condition to an issue during the supplier manufacturing process.Complaint information will continue to be monitored and actions assigned as required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17430988
MDR Text Key320115506
Report Number3012307300-2023-07648
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC37101329-NLJ
Device Lot Number4230619
Was Device Available for Evaluation? No
Date Returned to Manufacturer07/12/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-