SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101329-NLJ |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 01/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a leak alarm occurred and use was discontinued.Patient involvement unknown.
|
|
Manufacturer Narrative
|
Other, month and year of event have been provided, day is unknown.Udi number is unknown, no information has been provided to date.510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Other text: h3, h6: event methods, evaluation, and conclusion codes: updated.Four photos of the device and one sample were received for investigation.The reported issue was confirmed during visual inspection, which confirmed there was a hole in the device breathing bag.A review of the manufacturing device history records (dhr) shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with the reported lot of products.The investigation attributed the root cause of this condition to an issue during the supplier manufacturing process.Complaint information will continue to be monitored and actions assigned as required.
|
|