EDWARDS LIFESCIENCES EDWARDS INSPIRIS RESILIA AORTIC VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 11500A
Device Problems Gradient Increase (1270); Device Stenosis (4066)
Patient Problems Dyspnea (1816); Fatigue (1849); Muscle Weakness (1967); Dizziness (2194); Insufficient Information (4580)
Event Date 07/06/2023
Event Type  Injury  
Manufacturer Narrative
H10: additional manufacturer narrative: surgical/percutaneous intervention is indicated or performed, or harm occurred due to the device, or there is a device malfunction that could cause or contribute to a serious injury.This event is considered a serious injury.The device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was reported that a patient with an 23mm 11500a valve underwent a valve-in-valve procedure after an implant duration of approximately of 7 years due to aortic stenosis.The procedure was performed with a 23mm 9750tfx transcatheter valve.
 
Event Description
It was reported that a patient with an 23mm 11500a valve underwent a valve-in-valve procedure after an implant duration of 6 years and 10 months due to aortic stenosis.The patient presented with shortness of breath, dizziness, weakness and fatigue.The procedure was performed with a 23mm 9750tfx transcatheter valve.The patient was stable post procedure and was discharged on pod #1.
 
Manufacturer Narrative
H10: additional manufacturer narrative: stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.The most likely cause is patient factors, including chronic kidney disease (ckd).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INSPIRIS RESILIA AORTIC VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key17435481
MDR Text Key320158623
Report Number2015691-2023-14952
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number11500A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/06/2023
Initial Date FDA Received08/01/2023
Supplement Dates Manufacturer Received08/18/2023
09/19/2023
Supplement Dates FDA Received09/12/2023
09/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Hospitalization; Required Intervention;
Patient Age65 YR
Patient SexMale
-
-