ABBOTT QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2023
Event Type  malfunction  
Event Description
During an implantation procedure, it was not possible to remove the guidewire from the left ventricular (lv) lead, which resulted in damage to the lead.A new lead was successfully implanted to resolved the event.The patient was stable.
 
Manufacturer Narrative
The reported event of difficulty removing the stylet that resulted in lead damage was confirmed.As received, a complete lead was returned in one piece.Stylet was not returned with the lead.Visual inspection of the lead found that the connector pin and crimp sleeve were pulled out of the connector assembly along with the inner coil which is consistent with procedural damage and the ptfe stylet coating was bunched up/clogged with the inner coil distal to the connector pin.The cause of the reported event was due to bunched up ptfe coating of the stylet inside the inner coil that prevented the removal of the stylet and excessive forces resulted in the connector pin and crimp sleeve to be pulled out of the connector assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17439462
MDR Text Key320195619
Report Number2017865-2023-37081
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeGR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberS000090357
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/14/2023
Initial Date FDA Received08/01/2023
Supplement Dates Manufacturer Received09/07/2023
Supplement Dates FDA Received09/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-