• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP 150 (S5 RP150); CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP 150 (S5 RP150); CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Decreased Pump Speed (1500)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/04/2023
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in (b)(6), germany.During maintenance and repair activity done by a livanova field service representative, in addition to encoder replacement, s5 system was flashed to a new firmware (r324).Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that rotary encoder of a s5 roller pump was replaced during maintenance activity since it gave an error message (error setpoint encoder).It cannot be excluded that pump speed was not adjustable.There was no patient involvement.
 
Manufacturer Narrative
As per follow-up communication, error message "shaft angle encoder fault pump 8" was displayed.It signals that the pump speed can no longer be adjusted correctly.As troubleshooting the shaft encoder (#96-414-506) was replaced.Functional test passed positively and unit was returned to service.Complaint database review revealed that unit was installed in 2018 and no further similar issues are reported.Based on the outcome of service activity, it cannot be ruled out that the most likely root cause of the reported event is a defective shaft encoder.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP 150 (S5 RP150)
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17441053
MDR Text Key320219439
Report Number9611109-2023-00373
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817900382
UDI-Public(01)04033817900382(11)180208
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/04/2023
Initial Date FDA Received08/02/2023
Supplement Dates Manufacturer Received08/24/2023
Supplement Dates FDA Received09/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-