• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET GMBH PAIR OF LEG PLATES, 4-PART; TABLE AND ATTACHMENTS, OPERATING-ROOM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET GMBH PAIR OF LEG PLATES, 4-PART; TABLE AND ATTACHMENTS, OPERATING-ROOM Back to Search Results
Model Number 113373BC
Device Problem Use of Device Problem (1670)
Patient Problem Laceration(s) (1946)
Event Date 07/12/2023
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.Event site name (e1): (b)(6).H3 other text: device not returned to manufacturer.
 
Event Description
On (b)(6) 2023 getinge became aware of an issue with one of our accessories ¿ 113373bc - pair of leg plates, 4-part.As it was stated, user's thumb was pinched while operating the release lever of the leg plate leading to a cut.According to the provided information, the cut was dressed and no medical intervention was required.We decided to report the issue based on the potential for serious injury if the situation, namely the user's body cut during use of the leg plate, was to reoccur.
 
Manufacturer Narrative
Getinge became aware of an issue with one of our accessories ¿ 113373bc - pair of leg plates, 4-part.As it was stated, the user's thumb was pinched while operating the release lever of the leg plate leading to a cut.According to the provided information, the cut was dressed and no medical intervention was required.The issue led to a delay in surgery of approximately 15 minutes, but there is no information that it led to prolonged anesthesia time for the patient.We decided to report the issue based on the potential for serious injury if the situation, namely the user's body cut during the use of the leg plate, was to reoccur.With the investigation performed it was concluded that upon the event occurrence, the device was being used for the patient¿s treatment, and thus was also directly involved with the reported incident.As no malfunction with the device was found, it was considered that the getinge device was up to the specification.A review of the received customer product complaints revealed that there were no serious injuries to a user nor to a patient or operator when this particular malfunction occurred.Comparing the number of claimed devices to the number of sold devices worldwide, we can assume that the failure ratio of the issue investigated herein is (b)(4) as there were three similar reportable customer product complaints related to the investigated issue.The device was inspected by the sales representative.There was no malfunction of the leg plates.The site was also visited by the getinge service technician and no intervention was necessary.The subject matter expert at the manufacturing site was contacted to establish the root cause of this issue.The sme assessed that the design implemented in the affected leg plates was developed with security against unwanted release and any pinching risk could not be identified.In the instruction for use (ifu 1333.73 en 07, page 11), the user is informed about danged resulting from improper handling.The user is also warned (ifu 1333.73 en 07, page 11), that when adjusting, moving or storing or table/ table top, the staff, the patient and the accessories are exposed to pinching and shearing hazards, particularly in the area around the joints at the head rest, back and leg plates.The user shall always ensure that no one can be subjected to pinching or shearing action or injured in any other way and that the accessories do not collide with any nearby objects.It is likely that the user utilized the leg plates disregarding safety notes and suggestions from the user manual.In summary, there was no malfunction of the investigated leg plates.Based on all available information the sme established that the root cause for the user's thumb being pinched, what led to a cut, while operating the release lever of the leg plate, was most likely related to the user error.We currently do not have any information that would warrant further action towards the device manufacturing or devices on the market, however as per our complaint handling processes will continue to monitor the customer experiences with the device for any future information.
 
Event Description
Manufacturer's reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAIR OF LEG PLATES, 4-PART
Type of Device
TABLE AND ATTACHMENTS, OPERATING-ROOM
Manufacturer (Section D)
MAQUET GMBH
kehler strasse 31
rastatt
GM 
Manufacturer (Section G)
MAQUET GMBH
kehler strasse 31
rastatt
GM  
Manufacturer Contact
holger ullrich
kehler strasse 31
rastatt 
GM  
MDR Report Key17442414
MDR Text Key320222557
Report Number8010652-2023-00067
Device Sequence Number1
Product Code BWN
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number113373BC
Device Catalogue Number113373BC
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received10/13/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/09/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-