• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM Back to Search Results
Model Number 9445-24
Device Problem Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2023
Event Type  malfunction  
Event Description
It was reported that signal loss over one hour occurred.Data was provided for evaluation.The allegation was undetermined.The probable cause could not be determined.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that signal loss over one hour occurred.The product was evaluated.An external visual inspection was performed and passed.Voltage test was performed and passed.Nordic bluetooth pairing test was performed and failed.Data was provided for evaluation.The allegation was confirmed.The probable cause was a defective transmitter.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key17443276
MDR Text Key320229634
Report Number3004753838-2023-151573
Device Sequence Number1
Product Code QDK
UDI-Device Identifier00386270001627
UDI-Public00386270001627
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
DEN170088
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/21/2024
Device Model Number9445-24
Device Catalogue NumberSTT-OE-002
Device Lot NumberP16502277
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/06/2023
Initial Date FDA Received08/02/2023
Supplement Dates Manufacturer Received08/18/2023
Supplement Dates FDA Received09/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient Weight111 KG
-
-