MAQUET CARDIOVASCULAR LLC HEMOPRO 2 W/VASOSHIELD; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4001
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Manufacturer Narrative
Trackwise id (b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure while attempting to cut and seal branches the vasoview hemopro 2 (w/vasoshield) would not activate/failed to deliver energy.The power supply was set at 3.They switched out the cord and the pigtail and it still was not working.They opened another kit with no issues to complete the case.No harm to the patient.
 
Manufacturer Narrative
Trackwise # (b)(4).The device was returned to the factory for evaluation on 07/21/2023.An investigation was conducted on 08/02/2023.A visual inspection was conducted.Signs of clinical use and evidence of blood were observed on the device.Evidence of charred material was also observed between the jaws.The heater wire and clear silicone insulation of the jaws were observed to be intact with no visual defects.An electrical evaluation was conducted.A pre-cautery test was performed per the direction for use (dfu) with a reference cable, adapter, and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.An activation and transection capability test was performed over four (4) repetitions using "max life test method stm2048073 rev aa.The device successfully transected tissue four (4) times.The jaws were gently cleaned of debris and char with a saline and gauze pad as indicated in the direction for use (cv000008979).A temperature and resistance test was conducted to evaluate the device function per hemopro 2 final test 90523436 rev w.The resistance value was measured at.68 ohms which is within specification.The device passed the temperature measurements test.The displayed temperature increased and turned green within the 2 second specified timeframe.The displayed temperature decreased once the toggle swivel was released.Based on the results of the evaluation, the reported failure "failure to deliver energy¿ was not confirmed.The lot # 3000295249 history record review was completed.There were ncmrs , rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.Specific actions for the failure modes are being maintained and documented under maquet's corrective and preventive action (capa) system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO 2 W/VASOSHIELD
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17443959
MDR Text Key320236898
Report Number2242352-2023-00651
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700451
UDI-Public00607567700451
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4001
Device Catalogue NumberVH-4001
Device Lot Number3000295249
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-