Catalog Number 966110 |
Device Problems
Detachment of Device or Device Component (2907); Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that at the beginning of the surgery, damage to the varus/valgus guide was observed.It did not affect the surgery at any time since the varus/valgus guide was used in the same way.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Additional information received: a.At the beginning of the surgery the guide is observed unarmed, of varus-valgus and the missing stylus.B.This was not an impediment to starting the surgery since i used the varus-valgus guide anyway, which worked for me unarmed.C.The doctor says that due to the missing stylus, he does not need it and that with the guide (angel wing) it is enough.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 (device code).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: according to the information received, "damage to the varus/valgus guide is observed¿.The product was not returned to depuy synthes, however photos were provided for review.See attachments ¿31_jul_2023_23_21_12_17674¿ and ¿source file - report de novedad clínica el rosario 420638¿.The photo investigation revealed that the t-slide lock was separated from the 966110, sp*2 femoral locating device.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the 966110, sp*2 femoral locating device would contribute to the complained device issue.Based on the investigation findings, potential cause can be traced to end of device life and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
|