The reported field event of lead connection issue was confirmed.The device above elective replacement indicator (eri) was received in one piece for analysis.Septum material was observed inside the left ventricular set screws hex cavity.This did not allow the set screw to be tightened and secured properly in the field, resulting in high lead impedance.The issue was consistent with having occurred during the procedure.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were normal with no anomalies found.
|