• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM E; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM E; PROSTHESIS, HIP Back to Search Results
Catalog Number 010000849
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/03/2023
Event Type  malfunction  
Event Description
It was reported that during the procedure, the liner would not assemble with the shell.The procedure was completed with another device.There was no harm or health consequences to the patient.Attempts have been made and no further information is available.
 
Manufacturer Narrative
(b)(4).G2: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.One g7 neutral e1 liner 32mm e item# 010000849 lot# 6742238 was returned and evaluated.Upon visual inspection the locking feature of the device is deformed and there are circular indentations to the top of the od.Event confirmed based on the evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.A definitely root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL E1 LINER 32MM E
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17461043
MDR Text Key320702149
Report Number0001825034-2023-01818
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304526358
UDI-Public(01)00880304526358(17)250210(10)6742238
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number010000849
Device Lot Number6742238
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/13/2023
Initial Date FDA Received08/04/2023
Supplement Dates Manufacturer Received11/13/2023
Supplement Dates FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN SHELL.
-
-