A low readings issue was reported with the adc device.Customer reported receiving unspecified low readings obtained on the adc sensor scan and, as a result, experienced "weakness," and "bad peripheral circulation" and was unable to self-treat.Customer presented at the hospital and after a hcp meter reading of 450 mg/dl was obtained in comparison to an adc sensor scan reading of approximately 60 mg/dl, the customer was administered subcutaneous insulin and an unspecified intravenous medication by a hcp for treatment of a diabetic ketoacidosis diagnosis.No further treatment was indicated.There was no report of death or permanent injury associated with this event.
|
A low readings issue was reported with the adc device.Customer reported receiving unspecified low readings obtained on the adc sensor scan and, as a result, experienced "weakness," and "bad peripheral circulation" and was unable to self-treat.Customer presented at the hospital and after a hcp meter reading of 450 mg/dl was obtained in comparison to an adc sensor scan reading of approximately 60 mg/dl, the customer was administered subcutaneous insulin and an unspecified intravenous medication by a hcp for treatment of a diabetic ketoacidosis diagnosis.No further treatment was indicated.There was no report of death or permanent injury associated with this event.
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned and a valid serial number has not been provided.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Clinical data was reviewed and confirmed that libre sensors continue to be safe, effective, and perform as intended in the field.Stability data for libre sensors was reviewed and showed no anomalies or non-conformances that could have lead to the complaint.A tripped trend review was conducted for the reported complaint and fs libre sensors, no trends were identified that would indicate any product related issues.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|