Based on the electronic device logbook, it was determined that the reported event was caused by an interruption of the internal can communication by the cpu in the vgc (ventilation and gas controller).In accordance with its safety concept, the unit responded with an emergency shutdown of the components and issued the appropriate alarms.In this case, during ventilation, there is an autonomous change to monitoring mode to ensure the possibility of manual ventilation using the emergency o2 dosage (with anesthetic gas).Ventilation as well as gas monitoring remains unaffected by this.Since the device has already exceeded the predefined expected service life of almost 21 years, no in-depth analysis of the printed circuit board was performed.However, it cannot be ruled out that the interruption of internal communication was caused by electromagnetic radiation or electrostatic discharge.
|