• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8603800
Device Problems Gas Output Problem (1266); Failure to Deliver (2338); Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
It was reported that the device failed during operation, no more settings were possible.
 
Event Description
It was reported that the device failed during operation, no more settings were possible.
 
Manufacturer Narrative
Based on the electronic device logbook, it was determined that the reported event was caused by an interruption of the internal can communication by the cpu in the vgc (ventilation and gas controller).In accordance with its safety concept, the unit responded with an emergency shutdown of the components and issued the appropriate alarms.In this case, during ventilation, there is an autonomous change to monitoring mode to ensure the possibility of manual ventilation using the emergency o2 dosage (with anesthetic gas).Ventilation as well as gas monitoring remains unaffected by this.Since the device has already exceeded the predefined expected service life of almost 21 years, no in-depth analysis of the printed circuit board was performed.However, it cannot be ruled out that the interruption of internal communication was caused by electromagnetic radiation or electrostatic discharge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key17467759
MDR Text Key320480637
Report Number9611500-2023-00277
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8603800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-