Model Number G447
Device Problems Failure to Capture (1081); High Capture Threshold (3266)
Patient Problem Syncope/Fainting (4411)
Event Date 07/25/2023
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited loss of capture (loc) and high capture thresholds on the left ventricular (lv) channel.It was noted that the patient had a much lower capture threshold at implant.The device recommended lv pacing only.It was noted that the lv thresholds increased.Thus, the output was increased.Additionally, the patient developed a complete heart block.The patient experienced a syncopal episode, which indicates that there was loc.A thresholds test was performed and revealed that the thresholds increased further.Technical services (ts) provided potential programming options.The device was reprogrammed.The device remains in use.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17469185
MDR Text Key320485228
Report Number2124215-2023-41744
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number295176
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age70 YR
Patient SexMale
-
-