• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4000
Device Problems Failure to Deliver Energy (1211); No Apparent Adverse Event (3189)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4) the device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that the vasoview hemopro 2 was not working.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, it was discovered that the vasoview hemopro 2 was not working.It started beeping and now they believe they engaged the safety fuse they forgot about.They did not let it rest for 30 seconds.No patient harm.
 
Manufacturer Narrative
Trackwise#: (b)(4).Updated sections: b4, b5, d10, e1(event site mail), g4, g7, h2, h3, h6, h10, h11.Corrected sections: h6--health effect ¿ clinical code corrected from "4580" to "4582" h6--medical device ¿ problem code corrected from "1211" to "3189".For the two lot #'s 3000323871 and 3000321482.There were no ncmrs, rework, or deviations documented for the last 3 lots shipped to the account.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.For the lot #3000322018.The lot # 3000322018 history record review was completed.There were ncmrs , rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.H3 other text : device discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17470633
MDR Text Key320706361
Report Number2242352-2023-00661
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/17/2023
Initial Date FDA Received08/07/2023
Supplement Dates Manufacturer Received08/18/2023
Supplement Dates FDA Received09/06/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
-
-