• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN LINER; PROTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN LINER; PROTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Naturally Worn (2988)
Patient Problem Failure of Implant (1924)
Event Date 07/12/2023
Event Type  Injury  
Event Description
It was reported that the patient had a revision surgery due to poly wear.It was reported that no additional information was available.
 
Manufacturer Narrative
(b)(4).G2: foreign: australia.Visual examination of the provided pictures identified the liner has damage and there is a hole worn through the liner.Complaint confirmed based on evaluation of provided image.Lot identification is necessary for review of device history records, lot identification was not provided.Complaint history review cannot be performed without product identification.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: minimal superolateral eccentric positioning of the femoral component with respect to the central portion of the acetabular cup.Although not confirmed radiographically, this can be related to polyethylene wear.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN LINER
Type of Device
PROTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17470899
MDR Text Key320508772
Report Number0001825034-2023-01828
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue NumberUNK LINER
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/13/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age84 YR
Patient SexFemale
Patient Weight71 KG
-
-