(b)(4).G2: foreign: australia.Visual examination of the provided pictures identified the liner has damage and there is a hole worn through the liner.Complaint confirmed based on evaluation of provided image.Lot identification is necessary for review of device history records, lot identification was not provided.Complaint history review cannot be performed without product identification.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: minimal superolateral eccentric positioning of the femoral component with respect to the central portion of the acetabular cup.Although not confirmed radiographically, this can be related to polyethylene wear.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|