H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.A physical investigation was not possible.The user report contains information regarding failure to advance.Photos were provided for review.One of the pictures as focused wrong and nothing could be seen.The other picture showed a bit coating off on the guidewire.The reported in the images damage/ coating peel of the guidewire would not cause failure to advice.Most probable cause was mechanical jam of the catheter that resulted in guidewire damage.A clear root cause could not be identified but a damaged guidewire represents a known inherent risk.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|