Catalog Number 0279401200 |
Device Problem
Continuous Firing (1123)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/30/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that the device continuously activated.
|
|
Manufacturer Narrative
|
Alleged failure: electrode won't stop working the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be a probe short due to a fluid ingress, or an internal leak due to a broken/damaged bond of the suction tubing.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that the device continuously activated.
|
|