• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KAVO DENTAL GMBH MASTERMATIC LUX M25 L; DENTAL ELECTRICAL HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KAVO DENTAL GMBH MASTERMATIC LUX M25 L; DENTAL ELECTRICAL HANDPIECE Back to Search Results
Model Number M25 L
Device Problem Dent in Material (2526)
Patient Problem Burn(s) (1757)
Event Date 07/06/2023
Event Type  malfunction  
Event Description
During a crown preparation to tooth #3 the instrument got hot and burned patient slightly on right corner of lower lip.No medical care necessary.
 
Manufacturer Narrative
The handpiece has been checked visually and functionally against the valid specification.It was found that the head had a dent at the push button.This caused the push button to stick in which caused unusual friction and hence heat up of the push button.A dent is the result of a strong external force, e.G.A drop of the handpiece.To avoid such issues the ifu contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: warning: hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, untypical warming or when the cutter or grinder cannot be held.Do not use further and notify service.Caution: burning hazard from hot instrument head or hot instruments cover.If the instrument overheats, burns may arise in the oral area.Never contact soft tissue with the instrument head or instrument cover the following guidelines must be observed to ensure save use of the electrically driven contra-angle handpieces: the service instructions for contra-angle handpieces must be precisely following when using kavo spray or quattrocare care systems.Before each use, the contra-angle handpiece must be checked for external damage.Before each use, perform a test run with the contra-angle handpiece, and watch for atypical heating and unusual noise and vibration.Immediately stop using contra-angle handpieces that act unusual.Never press the pushbutton during operation.This also includes lifting the cheek or tongue.To ensure proper function, the medical device must be set up according to the reprocessing methods described in the kavo instructions for use, and the care products and care systems described therein must be used.Kavo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device.This service interval depends on the frequency of use and should be adjusted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERMATIC LUX M25 L
Type of Device
DENTAL ELECTRICAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM  88400
Manufacturer (Section G)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM   88400
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach / riss, 88400
GM   88400
MDR Report Key17478364
MDR Text Key320619001
Report Number3003637274-2023-00016
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Dentist
Type of Report Initial
Report Date 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM25 L
Device Catalogue Number1.009.3630
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/14/2023
Initial Date FDA Received08/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-