Catalog Number CS-27702-E
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).
 
Event Description
It was reported the physician found foreign material inside the arrow raulerson syringe when the physician withdrew the spring wire guide from the syringe.The issue was detected prior to patient use.No patient involvement with the device was reported.
 
Manufacturer Narrative
(b)(4).The customer report of a foreign material observed was confirmed by investigation of the returned sample.The customer supplied photos and video confirmed that there were foreign particulates coming from inside the arrow raulerson syringe (ars) valve.The customer returned an opened cvc kit including an introducer needle, guide wire assembly, and arrow raulerson syringe (ars) for evaluation.It was observed that the guide wire was advanced through the ars and there was a kink where the guide wire meets the proximal opening of the ars.The customer was contacted and responded that they were not aware of the kink.Visual inspection revealed that there were no signs of use observed.Further visual analysis of the guide wire assembly and ars revealed foreign particulate on the straightener tube and within the ars handle.Microscopic examination confirmed the foreign material inside the ars handle and straightener tube.Further analysis revealed that there was a hole in the internal valves of the ars.As per r & d engineering and manufacturing unit, a positive confirmation of the material could not be made based on the appearance of the substance and the small quantity p resent within the straightener tube and ars handle.Additionally, it could not be confirmed when the particulate was introduced into the kit.A device history record review was performed, and no relevant findings were identified.Based on the appearance of the sub stance and the small quantity present within the components, a positive confirmation of the material could not be made.Additionally, it cannot be confirmed when the particulate was introduced into the kit.Based on the sample received, comments from r & d, and the complaint investigation, the potential root cause cannot be determined.Teleflex will continue to monitor and trend for complaints of this nature.
 
Event Description
It was reported the physician found foreign material inside the arrow raulerson syringe when the physician withdrew the spring wire guide from the syringe.The issue was detected prior to patient use.No patient involvement with the device was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 2-LUMEN 7 FR X 20 CM
Type of Device
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key17481612
MDR Text Key320863713
Report Number3006425876-2023-00723
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00801902102362
UDI-Public00801902102362
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K900263
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCS-27702-E
Device Lot Number71F22H1871
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/12/2023
Initial Date FDA Received08/08/2023
Supplement Dates Manufacturer Received08/30/2023
Supplement Dates FDA Received09/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-