MAUDE Adverse Event Report: VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number PN-004 075
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Paralysis (1997)
Event Date 06/01/2022
Event Type  Injury  
Event Description
Information received on 7/27/2023 indicated an adverse event (ae) of paresis was found during a monitoring visit on (b)(6) 2022.This was not considered a procedure/device related adverse event because the paresis may have been present prior to the last ablation.Furthermore, the paresis is on the right side and no ablation was done on the right side of the heart (only left sided ablation).Therefore, it is unclear whether this ae is at all related to the ablation.However, in the letter from the consulted pulmonary physician, it was stated that the paresis was caused by the ablation.
 
Manufacturer Narrative
Additional information: g3, h2, h3.The results of the investigation are inconclusive since the device was not returned for analysis.In addition, no log files were received.Based on the information received, the cause of the reported paresis remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX  32470
Manufacturer (Section G)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17481818
MDR Text Key320659489
Report Number9680001-2023-00024
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPN-004 075
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/27/2023
Initial Date FDA Received08/08/2023
Supplement Dates Manufacturer Received08/14/2023
Supplement Dates FDA Received08/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient Weight84 KG
-
-