Catalog Number 826633 |
Device Problem
No Device Output (1435)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported: "five hours after the microsensor was implanted into the patient, there was no icp readings showed on the icp monitor.Then the physician used this icp monitor to connect with microsensor of other patient, it could show the icp readings normally.Therefore, the problem was with the microsensor.Then, the physician changed with another of the same device.".
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
The microsensor (id 826633) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
|