• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD ICP SENSOR CATHETER KIT; ICP MICROSENSORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD ICP SENSOR CATHETER KIT; ICP MICROSENSORS Back to Search Results
Catalog Number 826633
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/24/2023
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported: "five hours after the microsensor was implanted into the patient, there was no icp readings showed on the icp monitor.Then the physician used this icp monitor to connect with microsensor of other patient, it could show the icp readings normally.Therefore, the problem was with the microsensor.Then, the physician changed with another of the same device.".
 
Event Description
N/a.
 
Manufacturer Narrative
The microsensor (id 826633) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICP SENSOR CATHETER KIT
Type of Device
ICP MICROSENSORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17483967
MDR Text Key321549836
Report Number3014334038-2023-00132
Device Sequence Number1
Product Code GWM
UDI-Device Identifier10381780520436
UDI-Public10381780520436
Combination Product (y/n)N
PMA/PMN Number
K991222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number826633
Device Lot Number6396430
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/03/2023
Date Device Manufactured08/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-