The reported events were lead dislodgement and stylet could not be removed.As received, a complete lead was returned in one piece.Visual inspection of the lead found the stylet stuck inside the lead.Stylet could not be removed due to inner coil bunched up with the blood/body fluids distal to the connector boot which is consistent with procedural damage.The cause of the reported event of stylet could not be removed was isolated to the inner coil bunched up with blood/fluids.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead did not find any anomalies with the exception of procedural damage.The full measured s-curve hump height was within specification.
|