Catalog Number 367873
Device Problem Contamination of Device Ingredient or Reagent (2901)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/17/2023
Event Type  malfunction  
Manufacturer Narrative
B3: date of event is unknown.The date received by manufacturer has been used for this field.H.6 investigation summary: bd had not received samples, but 1 photo was provided for investigation.Evaluation of the photo indicated a tube with larger than normal droplets of additive on the tube wall.Due to this, the additive has yellowed slightly on irradiation.This is recognized to be an aesthetic defect with no impact on the efficacy of the tube.Additionally, 100 retained samples were visually inspected, and no additive abnormality was found.Bd was able to duplicate and confirm the customers¿ failure mode from the photo received.Bd was not able to identify a root cause for the indicated failure mode.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is able to be confirmed for the indicated failure mode.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends." h3 other text : see h.10.
 
Event Description
It was reported that while using the bd vacutainer® k2e 10.8mg plus blood collection tubes that there was foreign matter in tube; biological and nonbiological.The following information was provided by the initial reporter: one of the sites have reported an issue with an individual bd purple top tube.The hcp took a new pack out to use and found one of the tubes was contaminated.Lot number 3093802.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2E 10.8MG PLUS BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17486909
MDR Text Key321528702
Report Number9617032-2023-01066
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number367873
Device Lot Number3093802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/17/2023
Initial Date FDA Received08/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-