It was reported that the implantable cardioverter defibrillator (icd) exhibited a code 1005 indicative of an open circuit condition detected during shock delivery.Out of range shock impedance measurements were noted.Low out of range impedance measurements were observed on this right atrial (ra) lead as well as noise that was oversensed and loss of capture (loc).It is planned to perform a lead revision and replace both leads as well as a device upgrade in the near future.Additional information was received that a revision procedure was performed and the brady and tachy therapy of the device were deactivated.The rv lead and this ra lead were capped and replaced.No additional adverse patient effects were reported.This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
|