It was reported that during a device changeout procedure, this left ventricular (lv) lead tested fine, pre-operatively.Once connected to the new can, there was no pacing or sensing on the lv electrogram (egm); the egm was flat without markers.This lv lead was reconnected to the old can for testing, and again there was no pacing or sensing.This lv lead was then connected to the psa with similar results: no pacing or sensing.The lv configuration in the new can was able to achieve some sensing but no capture once reprogrammed ring to right ventride (rv).It was suspected that this lv lead had changed in some way.The implant status of this lv lead was unclear, however it was noted that the physician was attempting to implant a new lead.No adverse patient effects were reported.This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
|