MICROVENTION, INC WEB SL 17 SINGLE LAYER; INTRASACCULAR FLOW DISRUPTION DEVICE Back to Search Results
Model Number W5-4.5-2
Device Problem Difficult or Delayed Separation (4044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/18/2023
Event Type  malfunction  
Event Description
It was reported: a web sl 4.5 x 2 was opened and inserted into via-17 catheter.This web was completely unproblematic to insert/advance into the via17.Placement also went well, but they had to use the detachment system several times until it was finally detached.The web device remains implanted.Patient status is "good".There is no allegation of defect against the via-17 catheter.
 
Manufacturer Narrative
The device in question remains implanted.The ancillary devices used were stated to be returned to the manufacturer for evaluation but have not yet been received.The difficulty in separation as described could not be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.A search for non-conformances associated with this part/lot number combination and the part/lot combinations of the ancillary devices used, did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the devices.
 
Manufacturer Narrative
Items returned: - n/a.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications; potential complications include but are not limited to the following: hematoma at the site of entry, aneurysm rupture, emboli, vessel perforation, parent artery occlusion, hemorrhage, ischemia, vasospasm, clot formation, device migration or misplacement, premature or difficult device detachment, non-detachment, incomplete aneurysm filling, revascularization, post-embolization syndrome, and neurological deficits including stroke and death.Warnings and precautions: the web aneurysm embolization system is intended for single use only.The detachment control device is intended to be used for one patient.Do not resterilize and/or reuse the device.Reuse and/or resterilization can increase risk of infection, cause a pyrogenic response or other life threatening complications.Reuse and/or resterilization can degrade product performance, leading to device malfunction.Dispose of all devices in accordance with applicable hospital, administrative and/or local government policy.Advance and retract the device slowly.Do not advance the delivery device with excessive force.Determine the cause of any unusual resistance.Remove the device if excessive friction is noted and check for damage.Do not rotate the delivery device during or after delivery of the embolization device.Rotating the device may result in damage or premature detachment.The embolization device cannot be detached with any other power source other than a microvention inc.Detachment control device.Ensure that at least two detachment control devices are available before initiating an embolization procedure.Procedure.Detachment of the device: 31.The detachment control device is pre-loaded with batteries and will activate when the delivery device is properly connected.32.Verify that the rhv is firmly locked around the delivery device before attaching the detachment control device to ensure that the embolization device does not move during the connection process.33.Ensure that the delivery device gold connectors are clean and free from blood or contrast.If necessary, wipe the connectors with sterile water and dry before connecting.34.Insert the proximal end of the delivery device into the detachment control device.When the delivery device is properly connected, the light will flash green and an intermittent tone will be heard.35.Verify the embolization device position before pressing the detachment button.36.Push the detachment button.During firing, the light should be solid green and the beep should be continuous.37.Verify detachment by first loosening the rhv valve, then pulling back slowly on the delivery device and verifying that there is not embolization device movement.If the embolization device does not detach, push the detachment button again.If the device is still not detached, obtain a new detachment control device and attempt detachment up to two additional times.If it does not detach, remove the delivery device.38.Verify the position of the embolization device angiographically through the guide catheter.39.Prior to removing the microcatheter from the treatment site, place an appropriately sized guidewire completely through the microcatheter lumen to ensure that no part of the embolization device remains within the microcatheter.The physical device was not available for evaluation to determine if a condition existed that would have caused or contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.This information may be updated if additional information is provided at a later date.
 
Event Description
No additional information was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL 17 SINGLE LAYER
Type of Device
INTRASACCULAR FLOW DISRUPTION DEVICE
Manufacturer (Section D)
MICROVENTION, INC
35 enterprise
aliso viejo CA 92656
Manufacturer (Section G)
MICROVENTION, INC
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key17495355
MDR Text Key320768324
Report Number2032493-2023-00890
Device Sequence Number1
Product Code OPR
UDI-Device Identifier00842429102053
UDI-Public(01)00842429102053(11)210715(17)260630(10)21071511H
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberW5-4.5-2
Device Lot Number21071511H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/09/2023
Supplement Dates Manufacturer Received09/26/2023
Supplement Dates FDA Received10/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
VIA-17-154-01, LOT. 0000157812.; VIA-17-154-90, LOT. 0000105642.
-
-