• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134701
Device Problems Material Puncture/Hole (1504); Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent an unspecified ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and post-procedure, the bwi product analysis lab identified a hole on the pebax.During ablation, physician tried to pull back ablation catheter for changing sheath.He realized that blood trapped in ablation catheter's tip.He swapped and irrigate but the blood was not out.Finally, he decided to change catheter.There were no issues when removal or withdrawing the catheter.The physician confirmed that they did not identify any catheter or pebax damage.No patient consequences were reported.The blood trapped on the catheter tip is not mdr-reportable.The hole in the pebax is mdr-reportable.
 
Manufacturer Narrative
The photo analysis and product investigation were completed.Device evaluation details: a picture was received for evaluation following biosense webster's procedures.According to pictures provided by the customer, reddish material (presumably blood) was observed inside the pebax; however, no external damages were observed on the device.The product was returned to biosense webster for evaluation.Bwi conducted a visual inspection and functional test of the returned device.Visual inspection was performed, and a hole was observed on the pebax surface with reddish material inside.The root cause of the damage on pebax cannot be related to the manufacturing process since there is evidence that the device was manufactured in accordance with documented specifications and procedures.However, it could be related to the handling of the device during the procedure, but this cannot be conclusively determined.A manufacturing record evaluation was performed for the finished device batch number 30850656l, and no internal actions were identified.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17496242
MDR Text Key320783070
Report Number2029046-2023-01748
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835009774
UDI-Public10846835009774
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/18/2023
Device Catalogue NumberD134701
Device Lot Number30850656L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2023
Initial Date Manufacturer Received 07/21/2023
Initial Date FDA Received08/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SL1 8.5FR SHEATH
-
-