The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
The reported event is confirmed cause unknown.When evaluating the sample, the separation could be located at the proximal pigtail.The separation looks similar to what is seen when the material is cut with a sharp object due to the angle of the separation with no stretching or discoloration of the material.The suture was not provided for evaluation.No other defects were evaluated.The sample passed all dimensional requirements.The od was measured at 0.0800" and 0.0794"using a laser micrometer.The tip id was measured using a 0.043" pin gauge.A 0.038" guidewire passed through the sample with no hesitation.The tube id where the separation occurred was measured using a 0.050" pin gauge.Although an exact root cause could not be determined a potential root cause could be material selection.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent.Care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation.The insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure.¿ h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|