ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, the guidewire was stuck on the lead and when pulling it out, the left ventricular (lv) lead was damaged.The lead was not used, and a new lead was implanted.The patient was stable throughout the procedure.
 
Manufacturer Narrative
The reported event of guidewire insertion/removal issue was confirmed.As received, a complete lead was returned.Visual inspection of the lead found no anomalies to the lead body.X-ray examination found the inner coil was bunched at the connector region consistent with procedural damage.Electrical testing was normal.The cause of the reported event was isolated to the coil being offset/bunched at the connector pin region.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17498502
MDR Text Key321069366
Report Number2017865-2023-37873
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000141221
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/01/2023
Initial Date FDA Received08/10/2023
Supplement Dates Manufacturer Received09/06/2023
Supplement Dates FDA Received09/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GALLANT; OPTISURE; TENDRIL
Patient Age67 YR
Patient SexMale
-
-