Model Number 0998-00-3023-53
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during preventative maintenance by getting field service engineer, the cs300 intra-aortic balloon pump (iabp) unit failed battery run test.There was no patient involvement.
 
Event Description
N/a.
 
Manufacturer Narrative
The field service engineer (fse) replaced the batteries.The fse then completed check out and the checklist on the unit.The unit passed all functional testing.The following investigation was performed by technician of the maquet failure analysis and testing dept.(fat) wayne, nj.The failure analysis and testing dept.Received (2) batteries with a reported unit failure of a failed battery runtime test at 90 minutes.The fat performed a visual inspection and found the part to be in good condition.The fat installed the parts into cs300 test fixture and tested the part to factory specifications per procedure and the cs300 service manual.The battery runtime test failed at 13 minutes.Fat was able to verify the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17503138
MDR Text Key320824406
Report Number2249723-2023-03632
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 08/03/2023
Initial Date FDA Received08/10/2023
Supplement Dates Manufacturer Received03/11/2024
Supplement Dates FDA Received03/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/28/2008
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-