DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during preventative maintenance by getting field service engineer, the cs300 intra-aortic balloon pump (iabp) unit failed battery run test.There was no patient involvement.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
The field service engineer (fse) replaced the batteries.The fse then completed check out and the checklist on the unit.The unit passed all functional testing.The following investigation was performed by technician of the maquet failure analysis and testing dept.(fat) wayne, nj.The failure analysis and testing dept.Received (2) batteries with a reported unit failure of a failed battery runtime test at 90 minutes.The fat performed a visual inspection and found the part to be in good condition.The fat installed the parts into cs300 test fixture and tested the part to factory specifications per procedure and the cs300 service manual.The battery runtime test failed at 13 minutes.Fat was able to verify the reported issue.
|
|