Model Number HL-90
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Other text: b3 date of event is unknown, no information received to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the over temperature alarm will not trigger when test button is press.No patient involvement.
 
Manufacturer Narrative
Device evaluation: one device was returned for investigation.Visual inspection noted the device was received with scratches and nicks on both the enclosure and the front cover.The water tank had silicone on the inside and the outside on all four sides.The quick connect was corroded, had the old style power switch connected to the printed circuit board (pcb), faded line cord and a discolored pole clamp.Functional testing confirmed the complaint.When depressing the "over temp" button it did not alarm.Root cause was attributed to a faulty membrane switch.Switch failure was attributed to wear from usage.Replaced membrane switch, pcb, power switch cable, power switch retainer, quick connect fitting, water tank cover, and line cord.Performed preventative maintenance and calibrated.Device passed all functional testing after the completed repairs.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17503272
MDR Text Key321096092
Report Number3012307300-2023-08021
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer07/25/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-