It was reported that the patient with this cardiac resynchronization therapy pacemaker (crt-p) system was evaluated in the emergency department.Pauses were observed on the monitor and the patient experienced lightheadedness during them.The health care professional (hcp) was not able to provide the length of these pauses to boston scientific technical services (ts).The hcp stated that an external pacemaker was placed on this patient.This left ventricular (lv) lead exhibited noise which was suspected to be the result of electromagnetic interference (emi) due to the external pacemaker placed on the patient.Ts confirmed the noise was not detected by the device.However, loss of capture (loc) on this lv lead was suspected.Ts advised the hcp to contact a field representative for further evaluation.No additional adverse patient effects were reported.At this time, this device system remains in service.
|