• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. STEALTH 360 ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. STEALTH 360 ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number PRD-SC30-MICRO
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 07/18/2023
Event Type  Injury  
Event Description
Following low speed treatments in a heavily calcified anterior tibial artery (at) and dorsalis pedis artery using a stealth 360 orbital atherectomy device (oad) and balloon angioplasty, slow flow was observed in the peroneal artery and posterior tibial artery.Balloon angioplasty was performed and nitroglycerine was administered into the distal vessels.Slow flow improved, however, was still present in the at at the end of the procedure.The patient was stable.
 
Manufacturer Narrative
The device history record for the reported oad could not be reviewed as the lot number was not provided.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.H6 health effect - clinical code 4581: slow/no flow.H6 investigation conclusion code 22: the stealth 360® peripheral orbital atherectomy system instructions for user manual states slow flow or no reflow phenomenon is a potential adverse event that may occur and/or require intervention with use of the system.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTH 360 ORBITAL ATHERECTOMY SYSTEM
Type of Device
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
tonia moskalets
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key17509536
MDR Text Key320876303
Report Number3004742232-2023-00204
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPRD-SC30-MICRO
Device Catalogue Number7-10059-0
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age75 YR
Patient SexFemale
Patient Weight82 KG
Patient RaceBlack Or African American
-
-