This report has been identified as b.Braun medical internal report number (b)(4).One (1) unused sample and one (1) used sample were provided for evaluation.The samples were visually evaluated and passed.The samples were then leak tested per specification with passing results, no leak was observed on the caresite valve.Based on the evaluation results, the reported defect was not confirmed.The returned product was functionally tested per specification and the reported defect was not able to be replicated or confirmed.The exact root cause could be determined at this time.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|