• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL EPIC¿ VALVE (MITRAL); HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EPIC¿ VALVE (MITRAL); HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Catalog Number E100-27M
Device Problem Patient-Device Incompatibility (2682)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/07/2023
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2020, a 27mm epic valve was successfully implanted.During an outpatient follow-up on (b)(6) 2023, pannus was observed on the epic valve.No patient symptoms were reported.The following day on (b)(6) 2023, the 27mm valve was emergently explanted, and a 25mm epic valve was implanted as a replacement.The explanted valve had pannus on it.The patient was reported as stable.
 
Manufacturer Narrative
The reported pannus on valve was confirmed.The investigation found fibrous pannus ingrowth on cusp 1.There was fibrous thickening on cusp 2 and outflow thrombus on all cusps.No inflammation or significant calcifications were present.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met specifications.The cause of reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC¿ VALVE (MITRAL)
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17510317
MDR Text Key320886586
Report Number2135147-2023-03506
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/15/2023
Device Catalogue NumberE100-27M
Device Lot Number6840707
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/18/2023
Initial Date FDA Received08/11/2023
Supplement Dates Manufacturer Received09/01/2023
Supplement Dates FDA Received09/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization; Life Threatening;
Patient Age66 YR
Patient SexMale
-
-