Model Number HL-90 |
Device Problems
Electrical /Electronic Property Problem (1198); Power Problem (3010)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Other text: b3: date of event and d4: udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the power switch was broken at the printed circuit board (pcb).Patient involvement is unknown.
|
|
Manufacturer Narrative
|
Evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection noted a broken power switch and printed circuit board (pcb).Functional testing confirmed the complaint, the connection between the power switch and pcb is damaged.Root cause was attributed to poor design and regular use of the power switch.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.This issue will continue to be monitored and further actions taken accordingly.
|
|