Model Number 3CX*FX15RE40C |
Device Problem
Partial Blockage (1065)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/18/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H6: component code: 525 - tube.Health effect - impact code: 2645 - no patient involvement.Health effect - clinical code: 4582 - no clinical, signs, symptoms or conditions.Medical device problem code: 1065 - partial blockage.Investigation findings: 3233 - results pending completion of investigation.Investigation conclusions: 11 - conclusion not yet available.
|
|
Event Description
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, there was a purge liquid outlet blockage in the sampling line.No patient involvement.The product was changed out.The surgery was completed successfully.
|
|
Event Description
|
Purge liquid outlet blockage in the sampling line.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on august 11, 2023.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date).E1 (initial reporter - address, city and zip code).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H4 (device manufacture date).H6 (identification of evaluation codes 11, 3331, 4114, 170, 25).Type of investigation #1: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #2: 3331 - analysis of production records.Type of investigation #3: 4114 - device not returned.Investigation finding: 170 - manufacturing process problem identified.Investigation conclusions:25 - cause traced to manufacturing.The actual sample was not returned; however, a video was provided.After review of the video, it was confirmed that there was a partial blockage in the pigtail line as there was a slow drip of fluid coming from the line.A representative retention sample from the product lot was previously tested, and no anomalies were noted during testing.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Manufacturer Narrative
|
The returned sample was inspected upon receipt to confirm that there was a partial blockage in the pigtail line as there was a slow drip of fluid coming from the line when di water was passed through.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Event Description
|
Purge liquid outlet blockage in the sampling line.
|
|