SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/18/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer¿s allegation was confirmed.The device evaluation found that the front panel was damaged and broken.No visual appearance issues were found.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus, the high flow insufflation unit had front panel damage: part of the one-digit pressure number and the two-digit flow rate number were not lit, making the number 9 look like number 7.The issue was found during preparation for use in a therapeutic laparoscopic colectomy.The procedure was completed with the same device with no delay.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned to olympus for evaluation and the customer's allegation was confirmed.Based on the results of the investigation, the phenomenon, ¿a part of the number disappears¿, occurred due to breakage of the front panel.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17513755
MDR Text Key321096688
Report Number3002808148-2023-08297
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/24/2023
Initial Date FDA Received08/11/2023
Supplement Dates Manufacturer Received10/04/2023
Supplement Dates FDA Received10/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-