• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G138
Device Problems Failure to Capture (1081); Pacing Problem (1439); Failure to Read Input Signal (1581); Under-Sensing (1661)
Patient Problems Chest Pain (1776); Unspecified Heart Problem (4454)
Event Date 07/27/2023
Event Type  malfunction  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that electrogram (egm) review was requested for this cardiac resynchronization therapy defibrillator (crt-d) due to patient symptoms of chest pain and palpitations.Upon review, there was atrial undersensing and suspected atrial non-capture.Additionally, small p-waves observed and stored pacemaker mediated tachycardia (pmt) episodes were stored.Stored atrial tachy response (atr) episodes suggest atrial undersensing was likely attributed to atrial flutter.Further more, stored pmt episodes appeared to be atrially driven, likely due to the atrial flutter as well.Technical services (ts) reviewed troubleshooting options.This crt-d system remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17514463
MDR Text Key321242680
Report Number2124215-2023-43124
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589188
UDI-Public00802526589188
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/04/2019
Device Model NumberG138
Device Catalogue NumberG138
Device Lot Number378147
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/28/2023
Initial Date FDA Received08/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age90 YR
Patient SexFemale
-
-