It was reported that electrogram (egm) review was requested for this cardiac resynchronization therapy defibrillator (crt-d) due to patient symptoms of chest pain and palpitations.Upon review, there was atrial undersensing and suspected atrial non-capture.Additionally, small p-waves observed and stored pacemaker mediated tachycardia (pmt) episodes were stored.Stored atrial tachy response (atr) episodes suggest atrial undersensing was likely attributed to atrial flutter.Further more, stored pmt episodes appeared to be atrially driven, likely due to the atrial flutter as well.Technical services (ts) reviewed troubleshooting options.This crt-d system remains in service.No additional adverse patient effects were reported.
|