Catalog Number 05336163190 |
Device Problems
High Test Results (2457); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The c502 module serial number was (b)(6).The last preventive maintenance visit was in (b)(6) 2023.The field service engineer (fse) identified broken hoses on the instrument and replaced them.Reagent pipettes, sample pipettes and syringe valves were replaced, instrument check results were within specification.Calibration and qc were acceptable.Repeatability testing was performed with a different analyzer and the results were comparable.The service actions resolved the issue.
|
|
Event Description
|
The initial reporter questioned the results for multiple patient samples tested for tina-quant hba1c gen.3 (hba1c) on a cobas 6000 c (501) module.Discrepant results were provided for 4 patient samples.Patient 1 initial result was 7.6%.The repeat result was 5.5%.On (b)(6) 2023 patient 2 initial result was 11.9%.This result was reported outside of the laboratory where it was questioned as the patient is not a diabetic.On (b)(6) 2023 the repeat result was 4.9%.On (b)(6) 2023 patient 3 initial result was 7.5%.The repeat result was 5.7 %.On (b)(6) 2023 patient 4 initial result was 7.4%.The repeat result was 5.6 %.The questionable results for patients 1, 3 and 4 were not reported outside of the laboratory.
|
|
Search Alerts/Recalls
|