During an initial implant procedure, it was difficult to secure the set screw with torque wrench.When measurements were taken on the device, a decrease in sensing and increase in pacing impedance were observed.A new device was successfully implanted to resolve the event.The patient was stable.
|
The reported event of setscrew, impedance, and sensing anomaly could not be confirmed.The device was above elective replacement indicator (eri) upon receipt.Analysis revealed the right ventricular setscrew was partially stripped and contained septum material inside their hex cavities.However, the setscrew wasn¿t fully stripped to the extent where it wasn¿t able to be engaged by the torque driver.The damage to the setscrew was consistent with having occurred during the procedure.
|