• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNKNOWN PERITONEAL DIALYSIS CATHETER; CATHETER, PERITONEAL, LONG-TERM INDWELLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNKNOWN PERITONEAL DIALYSIS CATHETER; CATHETER, PERITONEAL, LONG-TERM INDWELLING Back to Search Results
Device Problem Positioning Problem (3009)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
The catheter used by the patient is not a fresenius device.The manufacturer of the catheter, and further product information, is unknown.(b)(4).This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERITONEAL DIALYSIS CATHETER
Type of Device
CATHETER, PERITONEAL, LONG-TERM INDWELLING
Manufacturer (Section D)
UNKNOWN
MDR Report Key17531729
MDR Text Key321430854
Report NumberMW5138642
Device Sequence Number1
Product Code FJS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other
Type of Report Initial
Report Date 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Patient Sequence Number1
-
-