EDWARDS LIFESCIENCES LLC AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Device Problems Retraction Problem (1536); Failure to Advance (2524); Difficult to Open or Close (2921)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Event Description
Medtronic received information that during the implant of a transcatheter bioprosthetic valve, the valve was deployed in the correct aortic position.A mild leak was reported and a balloon aortic valvuloplasty (bav) was performed with a 25 mm edward's balloon.The balloon was delivered through one of the valve struts and caused the valve to move up.The balloon became lodged in the valve and could not be released.Several attempts were made to release the valve, however, was unsuccessful.The patient had open heart surgery to remove the valve and balloon and a surgical valve was implanted.No further adverse patient effects were reported.(b)(4).This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
MDR Report Key17534296
MDR Text Key321138601
Report NumberMW5141199
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other
Type of Report Initial
Report Date 05/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/12/2023